Installation of a Lonestar standalone System

Installation of a Lonestar System.

 

This document details the installation of the hardware for a standalone Lonestar system.  The software used to control the Lonestar system is detailed in the document “Lonestar Software for Standalone Systems”.

 

Once the Risk Assessments have been performed and a suitable site for the location of the Lonestar has been determined, remove the Lonestar from its peli-case.  Place the Lonestar system on a solid, level surface capable of supporting its weight.  Remove the Cole Palmer filter and other equipment from the packaging and check against the packing list to be sure that everything has been despatched.

 

Typically, standalone Lonestar units are used in conjunction with a pump to draw the sample through the inlet.  Certain pumps can be controlled via the Lonestar software; please consult the Owlstone Support staff for compatible pumps and to give advice on required specifications.  Care must be taken with the choice of pumps as explosive mixtures may occur, so an explosive-proof pump must be purchased if required, depending upon the sample type.

 Cole_Palmer.jpg

 

 

The Cole Palmer inlet filter supplied with the Lonestar must be fitted to the inlet of the Lonestar.  This is to prevent particles >1µm in diameter from entering the Lonestar, possibly causing damage to the FAIMS chip.

 

 

 

 Pump_plus_Lonestar.jpg

 

 

The pump inlet is connected to the Lonestar exhaust fitting, so the sample gas will be drawn through the Lonestar and subsequently through the pump. 

 

 

 

 Variable_pump.png

 

 

Connect the pump to a variable voltage power supply.

 

This is the simplest installation of the standalone Lonestar system.  The flowpath is summarised below with a pump drawing sample air only, with no dilution of the sample flow with cleaned ambient air.  The Lonestar scrubber must be blanked to prevent ingress of ambient air into the system.  A total of 2.5Lpm used for example purposes only.

 

 Flowpath_full.png

 Cap_scrubber.png

 

Once the hardware has been installed, turn on the Lonestar system.  The Lonestar system starts in Analyser Mode by default, but this mode of software operation is for developed methods rather than for method development.  Press the Advanced button to view the Lonestar parameters.

 

 Advanced_screenshot.png

 

The Purge must be set ON to align the Lonestar internal flowpath for use with a pump.

 

 Purge_on_screenshot.png

 

If, in this setup, the concentration of the analyte being monitored is too high at the Lonestar (loss of RIP suggesting system saturation has occurred), then a restriction can be placed before the inlet filter.  This restriction may be a needle valve, a rotameter, a mass-flow controller or a fixed orifice.  In the example below, a needle valve is shown.  The Lonestar scrubber must not be blanked in this setup to allow the scrubber to clean ambient air, which dilutes the sample flow.

 Flowpath_restricted.png

 Uncap_scrubber.png

 

 

 

This document is available to download as a .pdf for printing, see the link below:

Was this article helpful?
0 out of 0 found this helpful
Have more questions? Submit a request

Comments

0 comments

Please sign in to leave a comment.